These 25 questions come from our CRCST course bank and are spread across the areas of the exam outline. Answer all 25, press the button at the bottom, and you will see your score by area, the correct answer to each question and the reason every other choice is wrong. No account is needed. The CRCST exam has 150 multiple-choice questions (125 scored and 25 unscored pretest items) with a 3-hour time limit at Prometric. HSPA reports pass or fail and does not publish a percentage passing score.
When scissors are tested with test material made for that purpose, a sharp pair should
Correct answer: C. Scissors are tested by cutting the test material made for that purpose, and a pair in good condition makes a clean, even cut along the blades and right out to the tips without folding, chewing, or tearing it. The tips do the finest work during surgery, so a pair that cuts only near the pivot is not acceptable. Scissors that fail are removed from service and sent for sharpening or repair per facility policy.
- A is not correct because folding the material between the blades is a classic sign of dull or misaligned blades.
- B is not correct because cutting only near the pivot shows the distal blades are dull or out of alignment, which is a failed test.
- D is not correct because a tear, even in a straight line, means the blades are not shearing cleanly, so the test is failed.
Key point: cut the material cleanly out to the tips
Reference: HSPA Sterile Processing Technical Manual, 9th ed. (2023)
Metzenbaum scissors cut the test material cleanly near the pivot but chew and fold it at the tips. What should the technician do?
Correct answer: C. Scissors must cut cleanly all the way to the tips, and chewing or folding at the distal end shows the edges are dull or misaligned there. The pair is removed from service, tagged with the reason, and sent for sharpening or repair per facility policy, and a working replacement is placed in the set if one is available. Technicians do not adjust or repair instruments themselves.
- A is not correct because the tips are the working part of dissecting scissors, so passing near the pivot does not make the pair usable.
- B is not correct because changing the test material to get a different result is not a valid test and does not fix the dull tips.
- D is not correct because adjusting the pivot is a repair task for a qualified repair service and can damage the scissors.
Key point: Remove them from service and tag them for sharpening
Reference: HSPA Sterile Processing Technical Manual, 9th ed. (2023)
An emergency cart comes back to central supply after a code. What must be done before it returns to service?
Correct answer: A. After a code, the full contents of the cart are checked against the standard content list, because staff may not remember every item that was used, opened, or moved. Missing and used items are replaced, expiration dates are checked, the cart is cleaned per policy, and a new numbered lock is applied and its number recorded. The cart then goes back to its location ready for the next emergency. A cart returned incomplete could leave staff without a needed item during a code.
- B is not correct because a new lock on an incomplete cart would falsely signal that it is fully stocked.
- C is not correct because staff in a code may not recall every item used or opened, so the whole cart is checked against the standard list.
- D is not correct because leaving the drawers for later sends an incomplete cart back toward service.
Key point: Restock it to the standard list, check dates, and apply a new numbered lock
Reference: HSPA Sterile Processing Technical Manual, 9th ed. (2023)
When a curette or osteotome is inspected, which finding means it must be sent for repair?
Correct answer: B. Cutting instruments such as curettes, osteotomes, chisels, and gouges are checked under lighted magnification along the full cutting edge. Nicks, burrs, flat spots, and dull areas reduce cutting performance, can tear tissue or gloves, and can shed metal fragments. Instruments with these defects are removed from service and sent for sharpening or repair.
- A is not correct because a smooth, evenly polished edge is what a sharp, undamaged cutting instrument should look like.
- C is not correct because a matte or satin finish is a normal manufactured finish chosen to reduce glare, not a defect.
- D is not correct because a manufacturer's name or catalog number etched on the shank is standard identification.
Key point: A nick, burr, or dull area on the cutting edge
Reference: HSPA Sterile Processing Technical Manual, 9th ed. (2023)
How often should insulated electrosurgical instruments be tested with an insulation tester?
Correct answer: C. Insulation can be damaged during any procedure or during cleaning and handling, and the defects that cause stray electrosurgical burns are often invisible. Testing every insulated instrument each time it is processed, after cleaning and before packaging, catches failures before the item returns to a patient. Failed instruments are removed from service.
- A is not correct because many insulation failures cannot be seen, which is why electrical testing is done on every instrument.
- B is not correct because an instrument can be used many times in a month, and each use can damage the insulation.
- D is not correct because waiting for a reported problem means a patient may already have been exposed to a defective instrument.
Key point: Each time they are processed, before packaging
Reference: HSPA Sterile Processing Technical Manual, 9th ed. (2023); manufacturer's instructions for use (IFU)
Detergent solution is transferred into a spray bottle that will stay at the sink for use by staff throughout the shift. Why must the bottle be labeled?
Correct answer: A. The OSHA Hazard Communication Standard requires that chemicals in the workplace be labeled so employees know what they are handling and the hazards involved. When a chemical is transferred into a secondary container that is not used immediately by the person who filled it, the container must carry a workplace label with the product identifier and hazard information. Unlabeled bottles can lead to the wrong chemical being used on instruments or to injury from an unknown substance. The SDS for the product remains the source of full hazard information.
- B is not correct because the labeling requirement is an OSHA rule about chemical identity and hazards.
- C is not correct because the workplace labeling rule for chemicals comes from OSHA.
- D is not correct because EPA registration applies to disinfectants and pesticides, not instrument detergents.
Key point: OSHA requires workplace containers to identify the chemical and its hazards
Reference: OSHA 29 CFR 1910.1200 (hazard communication)
Over several completed washer cycles, a technician notices that the level in the detergent container has not dropped at all. What is the MOST likely concern?
Correct answer: A. A washer that completes its cycles while the chemical level stays the same is probably not delivering detergent, which can result from an empty or kinked feed line, a failed pump, a loose connection, or an air lock. The cycle may finish without an alarm, but the load was washed with water alone and may not be clean. The technician should stop using the washer, report it, and have the chemical delivery checked according to the operator manual. Loads run during the period in question are evaluated per facility policy, and a cleaning verification test is run before the washer returns to use.
- B is not correct because expiration does not stop the level from dropping when the pump draws product.
- C is not correct because each cycle uses detergent, so the level should drop with use.
- D is not correct because a washer that doses correctly always draws down the supply.
Key point: The chemical may not be feeding, so the loads may not be clean
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities); manufacturer's instructions for use (IFU)
While checking a loaded cart before the cycle, a technician finds two trays on the cart that are not listed on the load record. What should be done?
Correct answer: D. The load record lists every item in the load along with the sterilizer, cycle number, date, and operator, and each package carries the matching load control number. If items are missing from the record, a recall would not find them and a patient could receive an item from a failed load without anyone knowing. The record is corrected before the cycle so it matches what is actually in the chamber. In computerized systems, this is done by scanning each tray onto the load.
- A is not correct because every package must carry a load control label.
- B is not correct because a record built after a failure may be incomplete and is no longer reliable.
- C is not correct because the record must list the items so they can be found during a recall.
Key point: Add them to the load record before the cycle is started
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities)
What is the main reason sterilization loads are documented and every package carries a load control number?
Correct answer: D. Load documentation and lot numbers exist so that any item can be traced to the cycle that processed it, and every item from a failed cycle can be found and recalled. Good records allow a targeted recall instead of a department wide one and support notification when items were already used. They are also evidence for surveyors that monitoring took place. Cost and productivity data are secondary uses of tracking systems.
- A is not correct because storage time is managed through event related sterility and stock rotation, not load control numbers.
- B is not correct because cost tracking is a secondary use and is not why ST79 calls for load records.
- C is not correct because productivity reports are a secondary use of tracking data and are not why load records are kept.
Key point: To allow traceability and recall after a failed cycle
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities)
Many facilities record the load control number of each implant or tray in the patient's record. What is the purpose of this practice?
Correct answer: D. Recording the load control number in the patient record links the patient to the sterilization cycle that processed the items used. If that cycle is later found to have failed, as with a positive BI, the facility can identify which patients were exposed and notify their physicians. This link is especially important for implants. Tracking systems often capture it by scanning the tray at the point of use.
- A is not correct because sterilizer selection is based on the IFU and availability, not surgeon preference.
- B is not correct because storage location is not tracked through the patient record.
- C is not correct because sterilization cycles are not billed to individual patients through the load number.
Key point: To identify exposed patients if the load later fails
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities); The Joint Commission
Where is the load control label placed on a wrapped package?
Correct answer: A. The load control label is applied to the outside of every package where it can be read without opening it, typically on the package label or on chemical indicator tape. Writing directly on wrap or the paper side of a pouch can damage the barrier, and ballpoint pens can puncture the material. Felt tip indelible markers or printed labels are used. A label inside the tray could not be read until the package was opened.
- B is not correct because the paper side of a pouch is never written on because ink and pressure can compromise it.
- C is not correct because writing on wrap can damage the barrier, and a ballpoint pen can puncture it.
- D is not correct because a label inside the package cannot be read before opening, which defeats its purpose for recall.
Key point: On the package label or the indicator tape
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities)
A load record lists the sterilizer, cycle number, date, and operator, but not the items in the load. What is the problem with this record?
Correct answer: B. A load record lists the contents of the load along with the sterilizer, cycle, date, operator, and monitoring results. The contents list is what lets the department identify and find every item from a failed cycle. Without it, staff cannot tell which items were in the load and may have to treat far more items as suspect. Computerized tracking systems build this list when each package is scanned onto the load.
- A is not correct because the operator is a standard part of the load record.
- C is not correct because BI results are added when a BI is run, but a missing contents list is the gap that blocks recall.
- D is not correct because the printout records cycle parameters and does not list the items in the load.
Key point: Items could not be identified for recall if the load failed
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities)
What does a clean tag attached to a piece of patient care equipment communicate?
Correct answer: C. A clean tag, or a clean cover or bag with a tag, tells the next user that the equipment has been cleaned, disinfected, and inspected and has not been used since. Tags usually carry the date and the initials of the technician who processed the unit, per facility policy. The tag also helps staff on the units tell clean equipment from used equipment at a glance. A unit with a missing or broken clean tag is treated as soiled and reprocessed.
- A is not correct because patient assignment is recorded in the tracking system, not shown by a clean tag.
- B is not correct because patient care equipment is cleaned and disinfected, not sterilized, and a clean tag makes no sterility claim.
- D is not correct because PM status is shown on the Biomedical Engineering PM label, a separate tag from the clean tag.
Key point: It was cleaned and disinfected and is ready for issue
Reference: HSPA Sterile Processing Technical Manual, 9th ed. (2023)
On a hinged clamp, which area is MOST likely to retain soil after cleaning?
Correct answer: C. Soil collects where surfaces meet or where there are grooves and crevices that brushes and water spray have trouble reaching. On hinged instruments this means the box lock, where the two halves join, and the serrations or teeth of the jaws. These areas get the closest look under lighted magnification, with the instrument opened so the inner surfaces can be seen.
- A is not correct because flat, smooth shank surfaces are easy for brushes and spray to reach and rarely hold residue.
- B is not correct because the outside of the finger rings is smooth and open, so soil is removed there more easily.
- D is not correct because the handle center has no joints or grooves compared with the box lock and jaws, so it is less likely to trap soil.
Key point: The box lock and the jaw serrations
Reference: HSPA Sterile Processing Technical Manual, 9th ed. (2023)
Which cleaning tool is appropriate for removing soil from the surfaces of stainless steel surgical instruments?
Correct answer: A. Instrument surfaces are protected by a thin passivation layer of chromium oxide, and many instruments have fine serrations, coatings, or markings that abrasives can damage. Soft-bristled brushes, lint-free cloths, and sponges made for instrument cleaning remove soil without scratching. Steel wool, scouring powders, and abrasive pads scratch the surface, remove the protective layer, and create sites for corrosion and soil to collect. The device IFU names any special tools a particular instrument requires.
- B is not correct because abrasive pads scratch instruments and are not approved for device cleaning.
- C is not correct because steel wool scratches instruments and can leave metal particles that cause rust.
- D is not correct because abrasives damage the surface finish and passivation layer.
Key point: Soft-bristled brush made for instrument cleaning
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities); HSPA Sterile Processing Technical Manual, 9th ed. (2023)
While setting up the sinks at the start of a shift, a technician finds a reusable channel brush with frayed, bent bristles. What should the technician do?
Correct answer: B. Damaged brushes clean poorly because bent or frayed bristles do not contact surfaces evenly, and loose bristles can break off and remain inside a lumen. Brushes are inspected as part of daily work area preparation, and any that are worn or damaged are discarded and replaced. The replacement must be the size and type the device IFU calls for. Keeping an adequate stock of brushes at the sinks prevents technicians from using damaged or wrong-sized tools.
- A is not correct because trimming changes the brush size and leaves damaged bristles that may shed.
- C is not correct because detergent does not repair a damaged brush.
- D is not correct because a damaged instrument brush is discarded rather than reassigned.
Key point: Discard it and replace it with an intact brush of the correct size
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities); HSPA Sterile Processing Technical Manual, 9th ed. (2023)
A surgical case cart is assembled using:
Correct answer: A. Case carts are built for one scheduled procedure. The surgeon's preference card lists the instruments, supplies, and special items the surgeon uses for that procedure, and the pick list generated from it tells the technician what to pull and in what quantities. The cart is checked against the pick list before it leaves, and missing items are reported. Keeping preference cards current reduces returned, unused items and missing items in the OR.
- B is not correct because exchange cart counts track routine unit usage, not a procedure's instrument and supply needs.
- C is not correct because load records document sterilization cycles and do not list what a case requires.
- D is not correct because PAR worksheets restock unit supply areas and do not reflect what a specific surgery needs.
Key point: The surgeon's preference card and the case pick list
Reference: HSPA Sterile Processing Technical Manual, 9th ed. (2023)
A technician finds damp spots on several wrapped packages on a shelf below an air supply vent in sterile storage. What is the BEST action?
Correct answer: A. Under event-related sterility, a package stays sterile until something compromises it, and moisture is one of those events. Damp wraps can carry microorganisms through the barrier, so the affected packages are removed, the contents are reprocessed, and the packaging is replaced. The technician reports the condition to the supervisor and facilities engineering because condensation near a vent points to an environmental problem that will keep damaging stock. The location is checked again after repair.
- B is not correct because removing visible moisture does not reverse contamination that may have wicked through the wrap.
- C is not correct because the packages are already compromised and the moisture source would remain unreported.
- D is not correct because drying does not restore a barrier that has already been wet.
Key point: Reprocess the damp packages and report the moisture source
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities)
How often should floors and horizontal work surfaces in SPD processing areas be cleaned?
Correct answer: B. ANSI/AAMI ST79 recommends that floors and horizontal work surfaces in processing and storage areas be cleaned at least daily. These surfaces collect dust, lint, and spatter that can settle onto instruments and packages. Other surfaces such as walls, shelving, and vents are cleaned on a regular schedule set by facility policy. A written cleaning schedule and record shows that the work was done.
- A is not correct because a week of dust and lint on work surfaces and floors is far more than daily cleaning allows.
- C is not correct because monthly is far too infrequent for surfaces used every day.
- D is not correct because dust and microorganisms can be present before soil can be seen.
Key point: At least daily
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities)
Why are damp or wet cleaning methods preferred over dry sweeping and dry dusting in SPD?
Correct answer: A. Dust and lint carry microorganisms and can settle on work surfaces, open trays, and wrapped packages. Dry sweeping and dry dusting stir these particles into the air instead of removing them. Damp mopping and damp wiping trap the particles in the cleaning cloth or mop so they leave the room. Sterile storage and clean work areas are kept as free of dust and lint as possible for this reason.
- B is not correct because damp cleaning still uses a cleaning or disinfecting product as policy requires.
- C is not correct because room temperature is set by the ventilation system, not by the cleaning method.
- D is not correct because dry dusting does not leave a residue; the concern is airborne dust.
Key point: Dry methods put dust and lint into the air, where it settles on packages
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities)
Which type of product is used for routine cleaning and disinfection of SPD environmental surfaces such as counters and floors?
Correct answer: C. Environmental surfaces such as floors, walls, and work counters contact only intact skin or no skin at all, so they are cleaned and disinfected with an EPA-registered hospital disinfectant. The label sets the dilution and the wet contact time that must be met for the product to work. High-level disinfectants are regulated by the FDA for semicritical medical devices and are not used on room surfaces. Staff follow the product label and the SDS for safe use.
- A is not correct because water alone does not disinfect a surface.
- B is not correct because those products are for semicritical devices, and their fumes and toxicity make them unsuitable for room surfaces.
- D is not correct because it is made for cleaning soiled instruments and does not disinfect environmental surfaces.
Key point: An EPA-registered hospital disinfectant used per its label
Reference: CDC Guideline for Disinfection and Sterilization in Healthcare Facilities (2008); ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities)
During sorting in decontamination, microsurgical instruments from a mixed set are separated mainly so that they are
Correct answer: C. Microsurgical instruments have fine tips, delicate jaws, and small hinges that bend or break when heavier instruments shift against them in a sink, basket, or washer. Sorting them out at the start of decontamination lets the tech clean them by their own IFU, often in protective holders, mats, or dedicated baskets with tips protected. Keeping each device's parts together during sorting also prevents loss. Long soaks are avoided because extended contact with solution can corrode or stain instruments.
- A is not correct because heavy retractors are exactly what damages fine microsurgical tips.
- B is not correct because prolonged soaking can corrode instruments, and prompt cleaning prevents soil from drying.
- D is not correct because the goal of separation is protection, and delaying cleaning lets soil dry.
Key point: kept away from heavy instruments that could damage their fine tips
Reference: HSPA Sterile Processing Technical Manual, 9th ed. (2023); Manufacturer's instructions for use (IFU)
Which arrangement of instruments in a washer basket is CORRECT?
Correct answer: B. Washer spray and rinse water must reach every surface, so hinged instruments are opened to expose box locks and jaw surfaces, and cups, basins, and other concave items face down or are tilted so they drain and do not hold water. Heavy items go on the bottom so they cannot fall on or crush lighter instruments. Delicate items are placed in their own baskets or holders rather than under heavier pieces. These loading rules come from washer and device IFUs and are the same ideas used when assembling sets for sterilization.
- A is not correct because nested basins block spray, and upturned basins hold water that interferes with rinsing and drying.
- C is not correct because closed box locks block cleaning, and heavy items on top can damage the instruments below.
- D is not correct because weight on delicate items bends or breaks them.
Key point: Hinged items opened, heavy items on the bottom, concave sides facing down
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities); Manufacturer's instructions for use (IFU), washer-disinfector
Which email to the OR scheduling desk about a tray that will be ready later than planned is MOST professional?
Correct answer: C. Professional electronic messages are brief, factual, and easy to act on. A clear subject line tells the reader what the message is about, and the body states the problem, what SPD is doing, and when the tray will be ready. Assigning blame in a wide distribution, writing in all capitals, or using unfamiliar abbreviations makes the message harder to use and reflects poorly on the department. Any patient information included is limited to what the reader needs.
- A is not correct because abbreviations the reader may not know cause confusion and errors.
- B is not correct because all capitals read as shouting and are harder to read.
- D is not correct because blaming coworkers in a wide email is unprofessional and does not help the OR plan.
Key point: A clear subject line, a brief factual update, and the expected ready time
Reference: HSPA Sterile Processing Technical Manual, 9th ed. (2023)
The surgery schedule lists a laparoscopic cholecystectomy. Which instrumentation does the case require?
Correct answer: A. The term breaks into laparo (abdominal wall), chole (bile), cyst (bladder or sac), and the suffix -ectomy (removal). Chole plus cyst names the gallbladder, the sac that stores bile, so the procedure is removal of the gallbladder through small incisions in the abdominal wall using a scope. The tech who reads the term correctly can pull the laparoscopic set and confirm the case cart against the preference card.
- B is not correct because laparoscopic means a scope-based approach, and -ectomy means removal, not cutting into.
- C is not correct because the suffix -ectomy means removal, not viewing.
- D is not correct because the root chole means bile, so the organ is the gallbladder, not the urinary bladder.
Key point: Laparoscopic instruments for removing the gallbladder
Reference: HSPA Sterile Processing Technical Manual, 9th ed. (2023)
Score by area
The full course has explained questions like these for every area of the exam, timed practice exams and a diagnostic that shows where to start.
Start your free 3-day trial See what the full course includes
For eligibility, fees and how the exam is scored, read our CRCST and CSPDT exam guide.
MyTest by Acheron Review is an independent exam preparation company and is not affiliated with, endorsed by or sponsored by HSPA or CBSPD.