These 25 questions come from our CBSPD CSPDT course bank and are spread across the areas of the exam outline. Answer all 25, press the button at the bottom, and you will see your score by area, the correct answer to each question and the reason every other choice is wrong. No account is needed. The CSPDT exam has 150 multiple-choice questions (125 scored and 25 unscored pretest items) with a 3-hour time limit at PSI test centers. CBSPD reports a scaled score, and a scaled score of 70 passes.
Metzenbaum scissors are used mainly to
Correct answer: A. Metzenbaum scissors have long handles and relatively short, fine blades, usually curved, and are made for dissecting and cutting delicate tissue deep in a wound. Using them on sutures, wire or tough tissue dulls and misaligns the blades quickly. For that reason they are kept separate from suture scissors on the back table and are checked for a clean cut at the tips during inspection.
- B is not correct because dressings are trimmed with bandage scissors, which have a blunt, flattened lower tip.
- C is not correct because sutures are cut with straight Mayo scissors or suture scissors, and wire with wire cutters, to protect fine dissecting blades.
- D is not correct because heavy fascia is cut with curved Mayo scissors, which have heavier blades.
Key point: dissect and cut delicate tissue
Reference: AORN Guidelines for Perioperative Practice (Guideline for Cleaning and Care of Surgical Instruments); manufacturer's instructions for use (IFU)
When taking off a splashed face shield at the end of a decontamination shift, which part should the technician handle?
Correct answer: C. The outside front of goggles and face shields is treated as contaminated during removal. The CDC example removes eye and face protection after gloves, handling it by the headband or ear pieces, which are the least contaminated parts. The device then goes into the designated container for reprocessing or into the waste if it is disposable. Hand hygiene follows, and is repeated if the hands touched the front of the shield.
- A is not correct because the front of the visor took the splashes and is the most contaminated surface.
- B is not correct because with gloves off, touching the contaminated front would soil bare hands.
- D is not correct because the lower edge sits in the splash zone and is considered contaminated.
Key point: The headband or the sides near the ears
Reference: CDC Guideline for Isolation Precautions (2007)
Which description fits a sharps container that meets the OSHA Bloodborne Pathogens Standard?
Correct answer: B. The Bloodborne Pathogens Standard requires sharps containers that are closable, puncture-resistant, leakproof on the sides and bottom, and labeled with the biohazard symbol or color-coded. They are kept upright and close to where sharps are found, and they are closed and replaced when the contents reach the fill line. Staff never reach into a sharps container or empty one by hand. In decontamination, the container sits within reach of the sink so sharps found in trays go straight in.
- A is not correct because cardboard is not puncture-resistant or leakproof.
- C is not correct because an open container cannot be closed for transport, and emptying sharps into a bag creates injury risk.
- D is not correct because emptying sharps by hand exposes staff to needlesticks and cuts.
Key point: Closable, puncture-resistant, leakproof and labeled
Reference: OSHA 29 CFR 1910.1030 (bloodborne pathogens)
Under the Bloodborne Pathogens Standard, which sharps-handling practice is prohibited?
Correct answer: A. Reaching into or manually compressing the contents of a sharps container puts the hand in direct line with needles and blades, so OSHA does not allow it. Sharps containers are closable, puncture resistant, leakproof on the sides and bottom, labeled or color coded, kept upright, and replaced routinely so they are not overfilled. When a container reaches its fill line, it is closed and replaced. Reusable sharps containers are opened, emptied and cleaned only in ways that do not expose workers to injury.
- B is not correct because using an instrument to pick up sharps is the recommended work practice.
- C is not correct because closing a container at the fill line is the expected practice.
- D is not correct because OSHA requires sharps containers to be kept upright during use.
Key point: Pushing the contents down into a sharps container by hand
Reference: OSHA 29 CFR 1910.1030 (bloodborne pathogens)
A laparoscopic set reaches decontamination with soil caked dry inside the jaws and channels, although the case ended only an hour ago. Which failure is the MOST likely cause?
Correct answer: C. Soil can dry on instruments surprisingly fast in the warm, ventilated air of an operating room. Point-of-use treatment, which keeps soil moist with an enzymatic product or a water-moistened towel and flushes lumens per IFU, is what keeps a short delay from turning into hardened soil. Dried soil is much harder to remove, raises the risk of biofilm, and lengthens the decontamination process. When dried soil keeps arriving, SPD reports it so the point-of-use process can be reviewed with the user department.
- A is not correct because the set has not been through the washer yet, so loading could not have caused the dried soil.
- B is not correct because the soil was already dry when the set arrived, so the problem happened before decontamination.
- D is not correct because a missing label is a safety problem, but it has no effect on how fast soil dries.
Key point: The instruments were not kept moist at point of use
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities)
Which of these is a unit of measure?
Correct answer: A. A unit of measure describes the package quantity in which an item is ordered, issued or counted, such as each, box or case. The order, the receiving check and the restock all depend on everyone using the same unit of measure. Lot identifies a production batch, PAR is a set stock quantity, and REF marks the manufacturer's catalog number. Confusing these on an order can bring in far more or far less product than was intended.
- B is not correct because REF is the symbol for the catalog number on a device label.
- C is not correct because PAR is the quantity a location restocks to, not the package size an item comes in.
- D is not correct because lot is a production batch code used for traceability and recalls, not a package quantity.
Key point: Case
Reference: facility policy; ISO 15223-1 (symbols for medical device labels)
Exam gloves come 100 each to a box and 10 boxes to a case. A clinic needs 2,300 gloves, and none are on hand. What is the smallest whole-case order that covers the need?
Correct answer: B. One case holds 10 boxes of 100, or 1,000 gloves. Two cases supply 2,000, which is short of 2,300, so the smallest whole-case order that covers the need is 3 cases, giving 3,000 gloves. Working the conversion step by step, each to box to case, avoids errors that come from guessing. When an item is bought by the case, any extra goes into stock and is rotated first in, first out.
- A is not correct because 2 cases provide only 2,000 gloves, which leaves the clinic 300 short.
- C is not correct because 23 is the number of boxes needed, not cases, and ordering 23 cases would bring in 23,000 gloves.
- D is not correct because 4 cases would cover the need but is more than the smallest order that does so.
Key point: 3 cases
Reference: facility policy
A clinic asked for 6 of an irrigation syringe expecting 6 boxes, but 6 single syringes arrived. The item master lists the syringe as issued by the each. What MOST likely caused the shortfall?
Correct answer: B. The system issued exactly what the order said: 6 of an item that is issued by the each. The clinic thought in boxes, but the order did not say boxes, and no one matched the quantity to the unit of measure. Checking the unit of measure on every line before an order or requisition is submitted prevents this kind of error. Where an item is often needed in larger amounts, a different unit of issue can be set up through the proper process.
- A is not correct because a high PAR would cause overstocking, not a delivery of single units.
- C is not correct because a recall would remove the product entirely, not deliver it in the wrong quantity.
- D is not correct because the right product arrived; only the quantity was different from what the clinic intended.
Key point: The quantity was not matched to the item's unit of measure
Reference: facility policy
While checking a finished tray before wrapping, a technician sees that four hemostats were placed in the tray clamped shut on the last ratchet tooth. What should be done?
Correct answer: D. Hinged instruments are sterilized open, with the ratchets disengaged, so steam or other sterilant can contact the jaw surfaces and the box lock. Keeping instruments locked during heating also places stress on the box lock and can contribute to cracks over time. A stringer or rack made for sterilization holds ring-handled instruments open and organized so they stay in position during handling.
- A is not correct because closed jaws keep the sterilant from reaching the surfaces pressed together and stress the joint during heating.
- B is not correct because even one engaged ratchet tooth keeps the jaws in contact and does not allow full sterilant contact.
- C is not correct because a towel does not open the jaws, and adding textile to a metal tray can also add condensate.
Key point: Release the ratchets and place the hemostats open on a stringer
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities); AORN Guidelines for Perioperative Practice (Guideline for Cleaning and Care of Surgical Instruments)
A spine loaner arrives from the vendor for tomorrow's case, and at assembly the scale shows the instruments and their case together at 27 lb (12.2 kg). How should the department handle it?
Correct answer: B. Loaner sets are processed under the same standards as facility-owned sets, and the 25 lb (11.4 kg) total weight limit applies to them as well. When a loaner case is over the limit, the department works with the vendor to divide the contents into additional trays in a way the vendor's IFU supports, and documents the change. Facility policy usually describes how overweight loaners are handled so the issue is resolved before the set is needed.
- A is not correct because loaners are cleaned, inspected, packaged and sterilized like any other set, and no exemption from the weight limit exists.
- C is not correct because an indicator weighs almost nothing, and every package needs an internal chemical indicator.
- D is not correct because absorbent material and longer drying do not correct an overweight set and are not a validated fix.
Key point: Divide it into lighter trays as the vendor IFU and policy allow
Reference: ANSI/AAMI ST77 (containment devices); ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities); manufacturer's instructions for use (IFU)
Which problem is MOST directly caused by a dense concentration of metal in one instrument tray?
Correct answer: D. During steam sterilization, metal absorbs heat as steam condenses on it, and a dense mass of metal produces a large amount of condensate. If the load cannot dry before it is removed, the result is a wet pack, which is considered contaminated and must be reprocessed. Spreading weight evenly, staying within the 25 lb (11.4 kg) limit and following the container and tray IFU reduce this risk.
- A is not correct because the immediate consequence of dense metal is moisture at the end of the cycle, and storage corrosion depends on other conditions.
- B is not correct because edge loss comes from use and handling, while dense trays affect heating and drying.
- C is not correct because a process indicator on the outside of the package responds to exposure and is not faded by tray density.
Key point: Excess condensate that leads to wet packs
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities); ANSI/AAMI ST77 (containment devices)
Transient flora on a technician's hands
Correct answer: D. Resident (normal) flora live on and in the body and are not fully removed by washing, while transient flora are picked up from patients, items and surfaces and sit on the skin's surface. Transient organisms are the ones most often spread by hands, and routine hand hygiene removes or kills them. That is why hand hygiene is performed after removing gloves and before handling clean items, even when gloves were worn.
- A is not correct because transient organisms cause infection through many portals of entry, not only the bloodstream.
- B is not correct because routine hand hygiene with an alcohol-based rub or soap and water removes transient flora.
- C is not correct because living in the deep layers of the skin describes resident flora.
Key point: are picked up by contact and removed by hand hygiene
Reference: CDC Guideline for Hand Hygiene in Health-Care Settings (2002)
When a routine steam biological indicator reads positive, who is notified according to typical facility policy?
Correct answer: D. A positive BI starts a defined failure procedure, and the people who run that procedure need to hear about it right away. The supervisor organizes the recall and the investigation, and infection prevention judges the risk to any patients who may already have received items from the affected loads. Other parties, such as the sterilizer service company or surgeons, are brought in as the policy and the findings require. The technician who reads the result documents it and does not wait for a repeat test before reporting.
- A is not correct because maintenance may be part of the response, but it does not replace notifying the supervisor who starts the recall.
- B is not correct because the outside service company may be called later, but facility staff who manage the recall must be told first.
- C is not correct because surgeons are notified as policy directs when affected items were used, not as a routine first step for every positive BI.
Key point: The SPD supervisor and infection prevention
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities); facility policy
A technician argues that nail extenders are fine to wear in the preparation area because gloves are worn for most tasks. What is the BEST response?
Correct answer: D. Artificial nails and nail extenders hold more microorganisms than natural nails, and those organisms remain after hand hygiene. They also puncture gloves and wrappers, and gloves are not worn for every task in the preparation area. For these reasons, AORN guidance and facility policies do not permit artificial nails or extenders for staff who handle instruments and supplies, and CDC hand hygiene guidance also advises against them. Natural nails are kept short and clean.
- A is not correct because frequent glove changes do not remove the organisms on the nails underneath or prevent tears.
- B is not correct because length does not change the extra microbial load that extenders carry.
- C is not correct because extenders are not acceptable in any processing area, including decontamination, where glove tears expose the hands to soil.
Key point: Extenders harbor microorganisms and can tear gloves, so they are not worn
Reference: CDC Guideline for Hand Hygiene in Health-Care Settings (2002); AORN Guidelines for Perioperative Practice (Guideline for Hand Hygiene)
A Thursday steam BI from Sterilizer 2 is positive. The last negative BI from that sterilizer was on Monday. Which loads are recalled?
Correct answer: A. A negative BI is the last biological evidence that the sterilizer was killing spores, so every load from that machine after the last negative result is in doubt. Load records list which items were in each of those loads, and the lot numbers on package labels let staff find them in storage, on carts and in user areas. Items from other sterilizers are not part of this recall unless the investigation finds a shared cause, such as the steam supply. Recalled items are reprocessed before use.
- B is not correct because other sterilizers have their own monitoring results and are not recalled unless a common cause is found.
- C is not correct because the failure may have started at any point after the last negative BI, so loads between Monday and Thursday are also suspect.
- D is not correct because all items from the affected loads are recalled, whether or not they are implants.
Key point: All loads from Sterilizer 2 processed after Monday's negative BI
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities)
Why is hand hygiene still performed after removing gloves that showed no visible tears?
Correct answer: B. Gloves reduce hand contamination but do not prevent all of it. Small holes that cannot be seen, and contamination of the hands while gloves are being removed, can leave microorganisms on the skin. Hand hygiene right after glove removal deals with both problems before the worker touches clean items or surfaces. Gloves never replace hand hygiene.
- A is not correct because hand hygiene products do not restore skin oils, and that is not why it is done.
- C is not correct because powder is a separate concern, and many gloves are powder-free; the main reason is contamination of the hands.
- D is not correct because any glove material can have unseen defects, so the rule applies to every type of glove.
Key point: Gloves can have unseen defects, and hands can be soiled during removal
Reference: CDC Guideline for Hand Hygiene in Health-Care Settings (2002)
On a weekend night shift, a technician reads a positive steam BI. The supervisor cannot be reached. What should the technician do?
Correct answer: B. A positive BI is reported as soon as it is read, and the department's written procedure names who to call when the usual supervisor is not available. The technician starts the steps the policy assigns, such as removing the sterilizer from use and holding affected items, and escalates to the next level of authority. Waiting until Monday would let suspect items keep moving to patients. The BI and its control are documented as found.
- A is not correct because throwing away a positive result hides a possible failure, and the procedure is triggered by the first positive result.
- C is not correct because delaying the report allows items from failed loads to be used in the meantime.
- D is not correct because the sterilizer comes out of use when the first positive result is read, pending the investigation.
Key point: Follow the positive BI policy and contact the next person in the chain of command
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities); facility policy
A new technician says, "If something is legal, it is acceptable in SPD." Which statement corrects this view?
Correct answer: A. Law sets the enforceable minimum that every employee must meet, while ethics and professional standards describe how a conscientious technician behaves beyond that minimum. Much of what keeps patients safe in SPD, such as following an IFU when no one is checking or reporting a wet pack that could easily be ignored, is not written into any statute. Voluntary standards such as ANSI/AAMI ST79 and facility policy fill that space, and surveyors hold the department to them. Most decisions in SPD are governed by standards, IFUs and facility policy rather than by statute.
- B is not correct because every person in the department, certified or not, is expected to act ethically and follow professional standards.
- C is not correct because ethics and professional standards reach conduct the law does not address, so the two are not identical.
- D is not correct because facility policy binds employees whether or not a law repeats it, and policy is how most standards are put into practice.
Key point: Law sets the minimum; professional ethics guide conduct beyond it
Reference: Facility policy; ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities)
At the end of a steam cycle, the printout shows the exposure temperature dipped below the selected setting for part of the cycle. The load holds two trays the OR needs within the hour. What is the correct action?
Correct answer: D. Physical monitors are reviewed at the end of every cycle, and the operator initials the printout to show the cycle met its parameters before anything is released. A temperature that falls below the selected setting means the load did not receive the exposure the cycle was designed to deliver, so nothing in it can be considered sterile. The technician quarantines the whole load and reports to the supervisor, who decides on reprocessing and whether the sterilizer needs service. Releasing part of a failed load because of OR demand is the kind of pressure the stop-and-report rule exists to resist.
- A is not correct because a Type 1 process indicator only shows that the package was exposed to the process, not that the cycle reached its parameters.
- B is not correct because a failed parameter affects every item in the load equally, so no part of it can be released.
- C is not correct because the printout is reviewed before release, not later, and a failed cycle may point to a sterilizer problem that affects the next load too.
Key point: Hold the entire load and notify the supervisor of the failed parameter
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities)
A hypothermia unit's hoses and connectors were taken apart for cleaning. Before the unit is tagged clean, what must the technician do?
Correct answer: D. Equipment leaves SPD ready for use, so parts removed for cleaning are reassembled the way the manufacturer's IFU shows, with connectors seated, caps in place and any required fluid or reservoir steps completed. A function check per the IFU, such as powering on, running the self-test and confirming the unit heats, cools or alarms as designed, then shows the reassembly worked. Only a device that passes is tagged clean and covered. A device that fails the check is tagged out of service and sent to Biomedical Engineering.
- A is not correct because loose parts can be lost or mismatched and the unit cannot be verified as working.
- B is not correct because routine reassembly and function checks per IFU are SPD tasks, while Biomed handles repair and PM.
- C is not correct because clinical staff expect ready-to-use equipment, and a missed connection may not be found until the unit is on a patient.
Key point: Reconnect the parts per IFU and confirm the unit runs correctly
Reference: manufacturer's instructions for use (IFU); facility policy
What is the main goal of traffic control in the SPD?
Correct answer: C. Traffic control limits who enters each area, what they wear, and the routes that people, carts, and supplies take. Every person who moves through a clean area brings in skin cells, lint, and microorganisms, and soiled items moving on the wrong route can contaminate clean ones. Signs, attire rules, designated routes, and closed doors put the policy into practice. The result protects clean and sterile items and helps hold area pressures.
- A is not correct because vendors may enter with proper attire per policy, and traffic control is about contamination.
- B is not correct because staffing levels are set separately and are not the purpose of traffic control.
- D is not correct because the shortest route can cross clean areas, which is exactly what traffic control prevents.
Key point: Limiting people and movement that carry contamination into clean areas
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities); AORN Guidelines for Perioperative Practice (Guideline for Surgical Attire)
Which steam condition is linked to wet packs?
Correct answer: A. Steam for sterilization should be saturated, carrying the right amount of moisture at its temperature. Wet steam carries extra water droplets into the chamber and leaves it on the load, which can produce wet packs even when loading is correct. Steam quality problems often come from the facility's steam supply and are investigated with engineering. Superheated steam is the opposite problem: too dry to transfer heat well.
- B is not correct because a passing air removal test shows that air was removed, not that the steam carries too much water.
- C is not correct because saturated steam at the correct temperature is the condition sterilization depends on.
- D is not correct because superheated steam is too dry and causes poor heat transfer, not wet packs.
Key point: Wet steam that carries excess water droplets
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities)
Three peel pouches from the same basket show beads of moisture under their film once the load reaches room temperature. How are they handled?
Correct answer: C. Moisture inside a package after cooling makes it a wet pack, and a wet package can let microorganisms wick through its barrier. All three pouches are repackaged and processed again. The technician also checks the likely causes: pouches crowded or flat, pouches placed under metal, a cart loaded too full, or a cycle problem. Finding the cause prevents the next set of wet pouches.
- A is not correct because moisture that dries later does not restore the barrier, and the pouches are already considered contaminated.
- B is not correct because every pouch with moisture is a wet pack, so all three are reprocessed.
- D is not correct because using wet items sooner does nothing to address contamination.
Key point: Reprocess the pouches and look for the cause in loading or the cycle
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities)
Years after a vaccine series, a person's immune system responds within days when exposed to the organism. Which cells make this fast response possible?
Correct answer: D. When B cells and T cells respond to an infection or a vaccine, some become long-lived memory cells. On a later exposure, these cells recognize the organism and launch a faster and stronger specific response than the first one, often before illness develops. This is the basis of long-lasting active immunity, including protection from the hepatitis B vaccine.
- A is not correct because platelets help blood clot and do not provide immune memory.
- B is not correct because red blood cells carry oxygen and are not part of the immune response.
- C is not correct because neutrophils are nonspecific phagocytes and do not keep a memory of past organisms.
Key point: Memory cells
Reference: CDC Principles of Epidemiology
A technician is assigned to manually clean orthopedic trays at the decontamination sink. Which attire meets the expected PPE for this task?
Correct answer: D. Manual cleaning creates splashes and sprays, so decontamination PPE protects the body, hands, face and hair. The expected ensemble is a fluid-resistant gown with sleeves, heavy-duty or task gloves with cuffs over the gown sleeves, a mask with goggles or a full face shield, and a head cover, with fluid-resistant shoe covers or boots when the floor is wet. Heavy-duty gloves resist cuts and punctures better than exam gloves. Missing any one barrier leaves a route for contaminated fluid to reach skin or mucous membranes.
- A is not correct because scrubs alone give no fluid barrier over the arms and body, and no head cover is listed.
- B is not correct because a cloth coat absorbs fluid, and thin exam gloves tear easily on sharp orthopedic instruments.
- C is not correct because ordinary eyeglasses leave the sides, top and bottom of the eyes open to splashes.
Key point: Fluid-resistant gown, task gloves, mask and goggles, head cover
Reference: ANSI/AAMI ST79:2017 (comprehensive guide to steam sterilization and sterility assurance in health care facilities); OSHA 29 CFR 1910.1030 (bloodborne pathogens)
Score by area
The full course has explained questions like these for every area of the exam, timed practice exams and a diagnostic that shows where to start.
Start your free 3-day trial See what the full course includes
For eligibility, fees and how the exam is scored, read our CRCST and CSPDT exam guide.
MyTest by Acheron Review is an independent exam preparation company and is not affiliated with, endorsed by or sponsored by HSPA or CBSPD.